Optimizing Volatile Organic Compound Determination by Static Headspace Sampling
Aplikace | | ZOEX/JSBInstrumentace
The measurement of volatile organic compounds in water is critical for environmental monitoring and regulatory compliance. Accurate quantification of VOCs at trace levels protects public health and informs remediation strategies. Static headspace sampling paired with GC/MS offers a streamlined workflow for VOC analysis, especially where European and Canadian regulations favor this approach over purge and trap.
This application note investigates automated static headspace sampling of over 50 VOCs in water using an on-line autosampler and GC/MS in SIM/Scan mode. The goal is to enhance sensitivity and meet or exceed USEPA Method 8260 performance criteria. Key objectives include method optimization for low-level detection, assessment of linearity, precision, accuracy, and method detection limits under static headspace conditions.
Sample preparation relied on salting out by adding 2 g of sodium chloride to 10 mL of standard solution. The optimized static headspace procedure was conducted using:
The headspace sampler was programmed to incubate samples at 60 °C for 20 minutes with agitation, followed by a 1 mL injection. The GC oven began at 45 °C (2 min hold) and ramped at 15 °C/min to 220 °C, with a total run time of 15 minutes. MS source, quadrupole and transfer line temperatures were set at 230 °C, 150 °C and 180 °C, respectively.
Calibration curves from 0.5–200 ppb for all target compounds demonstrated linearity with curve %RSD well under 15 %. Method detection limits met USEPA 8260 requirements. Seven replicate low-level standards yielded precision below 6 % RSD and recoveries averaging 101 %. Static headspace showed advantages such as absence of active trap sites, extended linear dynamic range and simplified autosampler integration. Limitations were addressed with optimized salt addition and SIM acquisition to improve detectability of poorly partitioning analytes.
Static headspace sampling with SIM/Scan acquisition offers:
This approach is well suited for routine monitoring of groundwater, drinking water and wastewater where method sensitivity and throughput are essential.
Advancements in autosampler design and MS acquisition modes will further lower detection limits and broaden compound coverage. Integration of high-throughput headspace-GC/MS workflows and improved data processing algorithms will enhance laboratory efficiency. Emerging microsampling techniques and miniaturized headspace systems may facilitate field-deployable VOC analysis.
The optimized static headspace GC/MS method achieved USEPA Method 8260 performance criteria for over 50 VOCs in water. Linearity, sensitivity, precision and accuracy were demonstrated, making static headspace a credible alternative to purge-and-trap for environmental VOC testing. The streamlined workflow reduces sample handling, maintenance demands and extends dynamic range.
1. USEPA Method 8260B, Revision 2, December 1996, Volatile Organic Compounds by GC/MS
GC/MSD, HeadSpace, GC/SQ
ZaměřeníŽivotní prostředí
VýrobceAgilent Technologies, EST Analytical
Souhrn
Importance of the Topic
The measurement of volatile organic compounds in water is critical for environmental monitoring and regulatory compliance. Accurate quantification of VOCs at trace levels protects public health and informs remediation strategies. Static headspace sampling paired with GC/MS offers a streamlined workflow for VOC analysis, especially where European and Canadian regulations favor this approach over purge and trap.
Aims and Overview of the Study
This application note investigates automated static headspace sampling of over 50 VOCs in water using an on-line autosampler and GC/MS in SIM/Scan mode. The goal is to enhance sensitivity and meet or exceed USEPA Method 8260 performance criteria. Key objectives include method optimization for low-level detection, assessment of linearity, precision, accuracy, and method detection limits under static headspace conditions.
Methodology and Instrumentation
Sample preparation relied on salting out by adding 2 g of sodium chloride to 10 mL of standard solution. The optimized static headspace procedure was conducted using:
- EST Analytical FLEX autosampler with 2.5 mL headspace syringe
- Agilent 7890 GC coupled to 5975 MS in SIM/Scan acquisition
- Restek Rxi-624Sil MS column (30 m × 0.25 mm I.D., 1.4 µm film)
- SKY splitless inlet liner (2 mm × 6.5 × 78.5 mm)
The headspace sampler was programmed to incubate samples at 60 °C for 20 minutes with agitation, followed by a 1 mL injection. The GC oven began at 45 °C (2 min hold) and ramped at 15 °C/min to 220 °C, with a total run time of 15 minutes. MS source, quadrupole and transfer line temperatures were set at 230 °C, 150 °C and 180 °C, respectively.
Main Results and Discussion
Calibration curves from 0.5–200 ppb for all target compounds demonstrated linearity with curve %RSD well under 15 %. Method detection limits met USEPA 8260 requirements. Seven replicate low-level standards yielded precision below 6 % RSD and recoveries averaging 101 %. Static headspace showed advantages such as absence of active trap sites, extended linear dynamic range and simplified autosampler integration. Limitations were addressed with optimized salt addition and SIM acquisition to improve detectability of poorly partitioning analytes.
Benefits and Practical Applications
Static headspace sampling with SIM/Scan acquisition offers:
- Regulatory compliance with European and Canadian protocols
- Simplified sample preparation without purge-and-trap complications
- Robust linearity and reproducible quantitation at trace levels
- Reduced maintenance by eliminating analytical traps and foam issues
This approach is well suited for routine monitoring of groundwater, drinking water and wastewater where method sensitivity and throughput are essential.
Future Trends and Applications
Advancements in autosampler design and MS acquisition modes will further lower detection limits and broaden compound coverage. Integration of high-throughput headspace-GC/MS workflows and improved data processing algorithms will enhance laboratory efficiency. Emerging microsampling techniques and miniaturized headspace systems may facilitate field-deployable VOC analysis.
Conclusion
The optimized static headspace GC/MS method achieved USEPA Method 8260 performance criteria for over 50 VOCs in water. Linearity, sensitivity, precision and accuracy were demonstrated, making static headspace a credible alternative to purge-and-trap for environmental VOC testing. The streamlined workflow reduces sample handling, maintenance demands and extends dynamic range.
References
1. USEPA Method 8260B, Revision 2, December 1996, Volatile Organic Compounds by GC/MS
Obsah byl automaticky vytvořen z originálního PDF dokumentu pomocí AI a může obsahovat nepřesnosti.
Podobná PDF
Optimizing of Volatile Organic Compound Determination by Static Headspace Sampling
|Agilent Technologies|Aplikace
JSB is an authorised partner of Optimizing of Volatile Organic Compound Determination by Static Headspace Sampling #134 Anne Jurek Introduction: Static headspace sampling has always been a viable option…
Klíčová slova
syringe, syringedelay, delayneedle, needleagitation, agitationbutylbenzene, butylbenzeneheadspace, headspacedepth, depthjsb, jsbfill, fillinjection, injectionspeed, speedoff, offsampling, samplingvocs, vocsflex
Optimizing Volatile Organic Compound Determination by Static Headspace Sampling 
|EST Analytical|Aplikace
Solid Phase Micro Extraction of Flavor Compounds in Beer ANNE JUREK Optimizing Volatile Organic Compound Determination by Static Headspace Sampling Application Note Environmental Author Abstract Anne Jurek Applications Chemist EST Analytical Cincinnati, OH Volatile contaminants in drinking, ground and wastewaters…
Klíčová slova
jurek, jurekanne, anneheadspace, headspacesyringe, syringesampling, samplingvocs, vocsdelay, delayneedle, needleagitation, agitationbutylbenzene, butylbenzenedepth, depthbeer, beerflavor, flavorstatic, staticfill
Static Headspace Sampling of Volatile Organic Compounds
|EST Analytical|Aplikace
Solid Phase Micro Extraction of Flavor Compounds in Beer ANNE JUREK Static Headspace Sampling of Volatile Organic Compounds Application Note Environmental Author Abstract Anne Jurek Applications Chemist EST Analytical Cincinnati, OH Volatile contaminants in drinking, ground and wastewaters are an…
Klíčová slova
jurek, jureksyringe, syringedelay, delayanne, anneneedle, needleagitation, agitationbutylbenzene, butylbenzenedepth, depthfill, fillinjection, injectionspeed, speedheadspace, headspacebeer, beeroff, offflavor
A Single Calibration Method for Water AND Soil Samples Performing EPA Method 8260
|Agilent Technologies|Aplikace
JSB is an authorised partner of #101 A Single Calibration Method for Water AND Soil Samples Performing EPA Method 8260 Anne Jurek Introducttion: The Unite ed States Environmental Protection…
Klíčová slova
centurion, centuriondesorb, desorbacetate, acetateest, estpurge, purgesampling, samplingtert, tertpling, plingtime, timebutylbenzene, butylbenzenebake, bakesamp, sampencon, enconwater, watermethyl