GC Troubleshooting Series Part One: Ghost Peaks
Technické články | 2009 | Agilent TechnologiesInstrumentace
Ghost peaks in gas chromatography signal underlying contamination and compromise analytical accuracy. Addressing these peaks is crucial for consistent, high-quality results in research and quality control environments.
This guide outlines a systematic framework to diagnose and eliminate ghost peaks in split/splitless inlets. It emphasizes a five-point path from sample preparation, carrier gas supply, inlet hardware, chromatographic column, to detector and data system interactions.
A stepwise diagnostic workflow includes:
Condensation tests reveal that increased peak intensity on a second blank indicates contamination upstream of the column. Backflash, caused by vapor exceeding liner volume, is mitigated by reducing injection volume, lowering inlet temperature, selecting larger liners, or pressure pulsing. Broad ghost peaks suggest carryover of high-boiling compounds, remediated by extended column bake-out (limited to 1–2 hours within thermal limits). Detailed inlet maintenance procedures ensure proper septum, liner, o-ring, and gold seal replacement without introducing new contaminants.
Implementing these troubleshooting and maintenance protocols enhances chromatographic reproducibility, reduces downtime, and maintains instrument integrity. Laboratories benefit from clearer baselines, reduced sample carryover, and predictable performance over extended operational periods.
Emerging developments include integrated diagnostic tools and software (e.g., LabAdvisor), advanced consumables with surface treatments to resist contamination, automated maintenance scheduling, and renewable filtration systems for carrier gas management. Such innovations will further streamline ghost-peak prevention and overall GC system reliability.
Ghost peaks serve as critical indicators of system contamination. A holistic approach combining sample preparation, controlled injection parameters, systematic inlet maintenance, and advanced consumables effectively eliminates these peaks and sustains high-quality GC analyses.
GC
ZaměřeníVýrobceAgilent Technologies
Souhrn
Significance of the topic
Ghost peaks in gas chromatography signal underlying contamination and compromise analytical accuracy. Addressing these peaks is crucial for consistent, high-quality results in research and quality control environments.
Objectives and overview
This guide outlines a systematic framework to diagnose and eliminate ghost peaks in split/splitless inlets. It emphasizes a five-point path from sample preparation, carrier gas supply, inlet hardware, chromatographic column, to detector and data system interactions.
Methodology and instrumentation
A stepwise diagnostic workflow includes:
- Blank and condensation tests to localize contamination before or after the column
- Evaluation of sample preparation materials, solvents, and SPE cartridges
- Inspection and maintenance of inlet components such as liners, septa, gold seals, and o-rings
- Control of injection parameters: volume, inlet temperature, residual vapor management (backflash)
- Routine bake-out procedures to remove semi-volatile residues
Main results and discussion
Condensation tests reveal that increased peak intensity on a second blank indicates contamination upstream of the column. Backflash, caused by vapor exceeding liner volume, is mitigated by reducing injection volume, lowering inlet temperature, selecting larger liners, or pressure pulsing. Broad ghost peaks suggest carryover of high-boiling compounds, remediated by extended column bake-out (limited to 1–2 hours within thermal limits). Detailed inlet maintenance procedures ensure proper septum, liner, o-ring, and gold seal replacement without introducing new contaminants.
Benefits and practical applications
Implementing these troubleshooting and maintenance protocols enhances chromatographic reproducibility, reduces downtime, and maintains instrument integrity. Laboratories benefit from clearer baselines, reduced sample carryover, and predictable performance over extended operational periods.
Future trends and applications
Emerging developments include integrated diagnostic tools and software (e.g., LabAdvisor), advanced consumables with surface treatments to resist contamination, automated maintenance scheduling, and renewable filtration systems for carrier gas management. Such innovations will further streamline ghost-peak prevention and overall GC system reliability.
Conclusion
Ghost peaks serve as critical indicators of system contamination. A holistic approach combining sample preparation, controlled injection parameters, systematic inlet maintenance, and advanced consumables effectively eliminates these peaks and sustains high-quality GC analyses.
Instrumentation Used
- Agilent GC system with split/splitless inlet and EPC carrier gas control
- Certified BTO septa and non-stick liners
- Gold-plated inlet seals and compatible o-rings
- Renewable gas filter cartridges for carrier gas purification
- Basic maintenance tools: tweezers, 1/4" and 1/2" open-end wrenches
Obsah byl automaticky vytvořen z originálního PDF dokumentu pomocí AI a může obsahovat nepřesnosti.
Podobná PDF
Better Connections - Agilent GC inlet resource guide
2021|Agilent Technologies|Příručky
Better Connections, Better Results with Your Agilent GC System GC inlet resource guide Here’s What’s Inside Introduction Septa Troubleshooting Ordering Information Ferrules Troubleshooting Ordering Information 2 3 4 4 7 8 9 13 Column Nuts Ordering Information Liners 14 15…
Klíčová slova
liner, linerproblem, probleminlet, inletferrules, ferrulescolumn, columnnormal, normalgraphite, graphitenut, nutliners, linersnuts, nutssepta, septaferrule, ferruleagilent, agilentwool, woolreplace
Don’t Fear the Ghost or Unexpected Peaks: Troubleshooting and Prevention
2019|Agilent Technologies|Prezentace
Don’t Fear the Ghost or Unexpected Peaks: Troubleshooting and Prevention Mark Sinnott Online Application Engineer Alexander Ucci Online Application Engineer October 10, 2019 1 October 10, 2019 Don’t Fear the Ghost or Unexpected Peaks: Troubleshooting and Prevention Agilent Restricted Exclusive…
Klíčová slova
restricted, restrictedagilent, agilentcontamination, contaminationsyringe, syringeinlet, inletghost, ghostsplit, splitjumper, jumperliner, linercleanup, cleanupemr, emrcolumn, columndirty, dirtyvent, ventvial
Don’t Fear the Ghost or Unexpected Peaks: Troubleshooting and Prevention
2019|Agilent Technologies|Prezentace
Don’t Fear the Ghost or Unexpected Peaks: Troubleshooting and Prevention Mark Sinnott Online Application Engineer Alexander Ucci Online Application Engineer October 10, 2019 1 October 10, 2019 Don’t Fear the Ghost or Unexpected Peaks: Troubleshooting and Prevention Agilent Restricted Exclusive…
Klíčová slova
restricted, restrictedagilent, agilentcontamination, contaminationsyringe, syringeinlet, inletghost, ghostsplit, splitjumper, jumperliner, linercleanup, cleanupemr, emrdirty, dirtycolumn, columnvial, vialvent
Agilent GC AND GC/MS - Catalog Consumables and Spare Parts
2016|Agilent Technologies|Ostatní
Buy online: www.agilent.com/chem/store Contact us: www.agilent.com/chem/contactus THE ESSENTIAL CHROMATOGRAPHY & SPECTROSCOPY CATALOG For more information This information is subject to change without notice. © Agilent Technologies, Inc. 2016 Printed in Canada March 8, 2016 5991-5213EN GC AND GC/MS Your Essential…
Klíčová slova
supplies, suppliesagilent, agilentparts, partscrosslab, crosslabdescription, descriptionoem, oemnut, nutinlet, inletliners, linerssepta, septacolumn, columnferrule, ferruleneedle, needlestick, stickassembly