Base neutrals
Aplikace | 2011 | Agilent TechnologiesInstrumentace
The reliable analysis of base neutral compounds is critical in environmental monitoring, regulatory compliance and water quality assessment. Many of these substances, such as nitrosamines and phthalates, pose health risks even at trace levels. Developing a fast, reproducible gas chromatography (GC) method enables laboratories to screen water and other environmental samples for a broad range of semi-volatile organic contaminants, facilitating timely decision-making and ensuring public safety.
This application note demonstrates a GC–flame ionization detector (FID) method for simultaneous separation and detection of 14 base neutral analytes in a single run. The goals were to achieve baseline resolution of structurally diverse compounds, minimize run time, and validate the approach for routine environmental analysis. The note highlights method parameters, chromatographic performance and peak identification.
A temperature-programmed GC method was developed to separate compounds ranging from low molecular weight nitrosamines to high molecular weight phthalates. Key method steps include:
The developed method achieved complete separation of all 14 base neutrals in under 30 minutes, demonstrating sharp peaks and good resolution even between structurally similar analytes. Key observations:
Advances in GC–MS coupling and high-throughput sample preparation are likely to further enhance sensitivity and selectivity for base neutrals. Potential directions include:
The described GC–FID method on an Agilent VF-200ms column provides a reliable, fast and reproducible approach for separating and detecting 14 common base neutral contaminants. Its balance of resolution, speed and robustness makes it a valuable tool for environmental laboratories tasked with routine monitoring of nitrosamines, phthalates and related compounds.
Agilent Technologies, Inc., "GC analysis of 14 base neutrals with an Agilent VF-200ms column," Application Note A02200, printed October 31, 2011, Agilent Technologies, USA.
GC, GC kolony, Spotřební materiál
ZaměřeníŽivotní prostředí
VýrobceAgilent Technologies
Souhrn
Significance of the topic
The reliable analysis of base neutral compounds is critical in environmental monitoring, regulatory compliance and water quality assessment. Many of these substances, such as nitrosamines and phthalates, pose health risks even at trace levels. Developing a fast, reproducible gas chromatography (GC) method enables laboratories to screen water and other environmental samples for a broad range of semi-volatile organic contaminants, facilitating timely decision-making and ensuring public safety.
Objectives and study overview
This application note demonstrates a GC–flame ionization detector (FID) method for simultaneous separation and detection of 14 base neutral analytes in a single run. The goals were to achieve baseline resolution of structurally diverse compounds, minimize run time, and validate the approach for routine environmental analysis. The note highlights method parameters, chromatographic performance and peak identification.
Methodology
A temperature-programmed GC method was developed to separate compounds ranging from low molecular weight nitrosamines to high molecular weight phthalates. Key method steps include:
- Injection of 1 µL sample in split mode (1:100) using methylene chloride solvent at 2,000 µg/mL concentration.
- Initial oven temperature set at 45 °C, ramped at 10 °C/min to 325 °C to elute late-eluting phthalates within a 28-minute window.
- Helium carrier gas flow maintained at approximately 1.0 mL/min under a 60 kPa pressure program.
- FID detection for sensitive measurement of carbon-containing base neutrals.
Used Instrumentation
- Gas chromatograph equipped with split/splitless inlet and FID.
- Agilent FactorFour VF-200ms column, 0.25 mm ID × 30 m length, 0.25 µm film thickness (Part no. CP8858).
- Carrier gas: helium at ~1.0 mL/min.
- Injector: split mode, split ratio 1:100.
- Detector: flame ionization detector.
Main results and discussion
The developed method achieved complete separation of all 14 base neutrals in under 30 minutes, demonstrating sharp peaks and good resolution even between structurally similar analytes. Key observations:
- Early-eluting nitrosamines (n-nitrosodimethylamine, n-nitrosodi-n-propylamine, n-nitrosodiphenylamine) were well resolved despite low boiling points.
- Chlorinated ethers and phenyl ethers exhibited distinct elution order, enabling clear differentiation of isomeric species.
- High molecular weight phthalates (di-n-octyl phthalate, bis-(2-ethylhexyl) phthalate) eluted without peak tailing at the high end of the temperature program.
Benefits and practical applications
- Comprehensive screening: One-run analysis covers nitrosamines, ethers, phenyl ethers and phthalates.
- High throughput: Short run time supports routine monitoring of multiple samples per day.
- Routine environmental compliance: Method aligns with regulatory needs for trace-level detection of base neutrals in water and soil extracts.
- Reproducibility: Well-defined temperature program and column choice ensure consistent results across laboratories.
Future trends and potential applications
Advances in GC–MS coupling and high-throughput sample preparation are likely to further enhance sensitivity and selectivity for base neutrals. Potential directions include:
- Integration with mass spectrometric detection for compound confirmation and lower detection limits.
- Automation of solid-phase extraction or large-volume injection to improve sample handling and trace-level quantitation.
- Application to broader matrices, such as wastewater, sediments and biological tissues.
- Use of fast GC columns with thin films and shorter lengths to reduce analysis time without sacrificing resolution.
Conclusion
The described GC–FID method on an Agilent VF-200ms column provides a reliable, fast and reproducible approach for separating and detecting 14 common base neutral contaminants. Its balance of resolution, speed and robustness makes it a valuable tool for environmental laboratories tasked with routine monitoring of nitrosamines, phthalates and related compounds.
Reference
Agilent Technologies, Inc., "GC analysis of 14 base neutrals with an Agilent VF-200ms column," Application Note A02200, printed October 31, 2011, Agilent Technologies, USA.
Obsah byl automaticky vytvořen z originálního PDF dokumentu pomocí AI a může obsahovat nepřesnosti.
Podobná PDF
Base neutrals
2011|Agilent Technologies|Aplikace
Base neutrals Application Note Environmental Authors Introduction Agilent Technologies, Inc. Agilent VF-1701ms columns separate 14 base neutrals in 37 minutes. Conditions Technique : GC-capillary Column : Agilent FactorFourVF-1701ms, 0.25 mm x 30 m fused silica (df= 0.25 μm) (Part no.…
Klíčová slova
neutrals, neutralscourtesy, courtesybase, baseprinted, printedinjector, injectorusa, usaauthors, authorsenvironmental, environmentalsize, sizetechnique, techniquecarrier, carrieridentification, identificationdetector, detectornote, notepeak
Base neutrals
2011|Agilent Technologies|Aplikace
Base neutrals Application Note Environmental Authors Introduction Agilent Technologies, Inc. Analysis of 14 base neutrals by gas chromatography using an Agilent FactorFour VF-1301ms column in less than 40 minutes. Conditions Technique : GC-capillary Column : Agilent FactorFour VF-1301ms, 0.25 mm…
Klíčová slova
neutrals, neutralscourtesy, courtesybase, baseprinted, printedinjector, injectorusa, usaauthors, authorsenvironmental, environmentaltechnique, techniquecarrier, carrieridentification, identificationdetector, detectornote, notepeak, peaktemperature
Base neutrals
2011|Agilent Technologies|Aplikace
Base neutrals Application Note Environmental Authors Introduction Agilent Technologies, Inc. The Agilent VF-17ms GC column separates 14 base neutrals in 28 minutes. Conditions Technique : GC Column : Agilent VF-17ms, 0.25 mm x 30 m fused silica (df = 0.25…
Klíčová slova
neutrals, neutralsbase, baseprinted, printedinjector, injectorusa, usaauthors, authorsenvironmental, environmentalsize, sizetechnique, techniquecarrier, carrieridentification, identificationdetector, detectornote, notepeak, peaktemperature
EPA 8270 Semi-volatile Analysis on an Agilent J&W FactorFour VF-5ms GC Column
2010|Agilent Technologies|Aplikace
EPA 8270 Semi-volatile Analysis on an Agilent J&W FactorFour VF-5ms GC Column Application Note Author Introduction Laura Provoost Agilent Technologies, Inc. U.S. Environmental Protection Agency (EPA) method 8270D, Semivolatile Organic Compounds By Gas Chromatography/Mass Spectrometry (GC/MS), is a challenging method.…
Klíčová slova
fluoranthene, fluoranthenebenzo, benzominutes, minuteschrysene, chryseneanthracene, anthracenebenz, benzpyrene, pyrenephthalate, phthalatebutyl, butylprovoost, provoostoctylphthalate, octylphthalatelaura, lauradiethylphthalate, diethylphthalatedimethylphthalate, dimethylphthalatenitrosodi