Precise Time-Scaling of Gas Chromatographic Methods Using Method Translation and Retention Time Locking
Technické články | 1998 | Agilent TechnologiesInstrumentace
GC
ZaměřeníVýrobceAgilent Technologies
Klíčová slovartl, aed, pressure, locking, method, psi, retention, points, calculated, translation, ret, calibration, time, pressures, scaling, times, scaled, original, cal, element, inlet, entered, using, translated, min, from, pesticide, column, selective, same, speed, measured, calculate, lock, new, scale, precise, library, would, locked, run, desired, hold, gain, file, used, match, detectors, corresponding, obtain
Podobná PDF
Precise Time-Scaling of Gas Chromatographic Methods Using Method Translation and Retention Time Locking
2006|Agilent Technologies|Aplikace
Precise Time-Scaling of Gas Chromatographic Methods Using Method Translation and Retention Time Locking Application Gas Chromatography May 1998 Authors Key Words B. D. Quimby, L. M. Blumberg, M. S. Klee, and P. L. Wylie Agilent Technologies, Inc. 2850 Centerville Road…
Klíčová slova
rtl, rtlaed, aedpressure, pressurelocking, lockingmethod, methodpsi, psiretention, retentionpoints, pointscalculated, calculatedtranslation, translationret, retcalibration, calibrationtime, timepressures, pressuresscaling
Excellent choices for environmental applications
2009|Agilent Technologies|Příručky
Solutions that meet your demands for: speed accuracy productivity Excellent choices for environmental applications Productivity Tools Applications > Return to Table of Contents > Search entire document Integrated Second Peristaltic Pump for Improved Sample Throughput and Reduced Matrix Effects Technical…
Klíčová slova
rtl, rtlpressure, pressuretime, timecolumn, columnaed, aedmethod, methodlocking, lockingagilent, agilentrtlock, rtlockicp, icpretention, retentionelement, elementcan, caninlet, inletflow
Excellent choices for environmental applications - Water
2009|Agilent Technologies|Příručky
Solutions that meet your demands for: speed accuracy productivity Excellent choices for environmental applications Water Volatiles Applications > Return to Table of Contents > Search entire document Ultra-Fast Total Petroleum Hydrocarbons (TPH) Analysis with Agilent Low Thermal Mass (LTM) GC…
Klíčová slova
icp, icpwater, wateragilent, agilenttime, timesample, samplewere, weremsd, msdanalysis, analysismass, masscolumn, columngas, gasusa, usamethod, methodmode, modeused
Retention Time Locking of Organochlorine Pesticides on an Agilent 8860 GC System Using the OpenLab Retention Time Locking Wizard
2019|Agilent Technologies|Aplikace
Application Note Food Testing Retention Time Locking of Organochlorine Pesticides on an Agilent 8860 GC System Using the OpenLab Retention Time Locking Wizard Author Ian Eisele Agilent Technologies, Inc. Abstract The retention time locking (RTL) wizard (a feature of Agilent…
Klíčová slova
locking, lockingbhc, bhcwizard, wizardrtl, rtlretention, retentionrelocked, relockedafter, afteropenlab, openlabpremaintenance, premaintenanceheptachlor, heptachlortrimming, trimminginlet, inletepoxide, epoxidetime, timeendrin