US EPA Method 524.3 with the Tekmar Lumin P & T Concentrator and Agilent 7890B GC and 5977A MSD System
Aplikace | 2017 | Teledyne LABSInstrumentace
Analyzing volatile organic compounds (VOCs) in drinking water at trace levels is essential for environmental monitoring and public health protection. Conventional purge and trap (P&T) techniques concentrate VOCs effectively but introduce substantial water vapor into gas chromatography–mass spectrometry (GC-MS) systems, leading to chromatographic interference, column degradation and frequent maintenance of the MS ion source.
This application note evaluates US EPA Method 524.3 performance when using the Teledyne Tekmar Lumin purge and trap concentrator equipped with a moisture control system (MCS) and the AQUATek 100 autosampler coupled to an Agilent 7890B GC and 5977A MSD. Key goals include:
Standard preparation involved Restek VOA MegaMix® calibration mixtures in deionized water with maleic and ascorbic acids. Calibration covered 0.2–50 ppb for target VOCs. Internal standards (1,4-difluorobenzene, chlorobenzene-d5, 1,4-dichlorobenzene-d4) and surrogate standards (MTBE-d3, 4-bromofluorobenzene, 1,2-dichlorobenzene-d4) ensured reliable quantification. A 1/X-weighted linear regression generated the calibration curves. Seven replicates at 0.5 ppb and seven at 5 ppb established MDLs, accuracy and precision.
Calibration across all 62 target VOCs exhibited coefficients of determination (r2) ≥ 0.995. MDLs ranged from 0.04 ppb to 0.18 ppb, meeting or exceeding Method 524.3 criteria. Accuracy (±20%) and precision (≤20% RSD) for 0.5 ppb and 5 ppb levels confirmed robust quantitation. Key observations included:
Enhanced trap cooling reduced desorption and reconditioning times, increasing sample throughput to over 90 samples in a 12 hour shift.
The integration of the Lumin P&T with MCS and AQUATek 100 delivers:
Emerging directions in VOC water analysis leveraging P&T technology include:
US EPA Method 524.3 analysis using the Tekmar Lumin P&T concentrator with moisture control and the AQUATek 100 autosampler, coupled to an Agilent 7890B GC and 5977A MSD, achieves accurate and precise quantification of trace VOCs in drinking water. The system demonstrates excellent calibration linearity, low MDLs and robust chromatography with minimal water interference, while enhancing throughput and reducing maintenance.
GC/MSD, Purge and Trap, GC/SQ
ZaměřeníŽivotní prostředí
VýrobceAgilent Technologies, Teledyne LABS
Souhrn
Significance of the Topic
Analyzing volatile organic compounds (VOCs) in drinking water at trace levels is essential for environmental monitoring and public health protection. Conventional purge and trap (P&T) techniques concentrate VOCs effectively but introduce substantial water vapor into gas chromatography–mass spectrometry (GC-MS) systems, leading to chromatographic interference, column degradation and frequent maintenance of the MS ion source.
Objectives and Study Overview
This application note evaluates US EPA Method 524.3 performance when using the Teledyne Tekmar Lumin purge and trap concentrator equipped with a moisture control system (MCS) and the AQUATek 100 autosampler coupled to an Agilent 7890B GC and 5977A MSD. Key goals include:
- Demonstrating reduced water transfer to the GC-MS without compromising VOC detection limits.
- Verifying calibration linearity, method detection limits (MDLs), accuracy and precision across a concentration range of 0.2 ppb to 50 ppb.
- Comparing sample throughput improvements via enhanced trap cooling and optimized cycle times.
Methodology and Instrumentation
Standard preparation involved Restek VOA MegaMix® calibration mixtures in deionized water with maleic and ascorbic acids. Calibration covered 0.2–50 ppb for target VOCs. Internal standards (1,4-difluorobenzene, chlorobenzene-d5, 1,4-dichlorobenzene-d4) and surrogate standards (MTBE-d3, 4-bromofluorobenzene, 1,2-dichlorobenzene-d4) ensured reliable quantification. A 1/X-weighted linear regression generated the calibration curves. Seven replicates at 0.5 ppb and seven at 5 ppb established MDLs, accuracy and precision.
- Tekmar Lumin P&T concentrator with MCS reduced moisture during desorption.
- AQUATek 100 autosampler provided reproducible sample handling and trap cooling enhancements.
- Agilent 7890B GC fitted with a 20 m × 0.18 mm DB-624 UI column (1 µm film) and operated under a temperature program from 35 °C to 225 °C.
- Agilent 5977A MSD scanned 35–260 amu with trace ion detection enabled.
Main Results and Discussion
Calibration across all 62 target VOCs exhibited coefficients of determination (r2) ≥ 0.995. MDLs ranged from 0.04 ppb to 0.18 ppb, meeting or exceeding Method 524.3 criteria. Accuracy (±20%) and precision (≤20% RSD) for 0.5 ppb and 5 ppb levels confirmed robust quantitation. Key observations included:
- First seven VOCs at 0.2 ppb showed clear characteristic ions with negligible water interference.
- A 5 ppb total ion chromatogram displayed sharp, well-resolved peaks for all analytes.
- The moisture control system effectively prevented excess water from eluting early compounds, improving peak shape and retention time stability.
Enhanced trap cooling reduced desorption and reconditioning times, increasing sample throughput to over 90 samples in a 12 hour shift.
Benefits and Practical Applications
The integration of the Lumin P&T with MCS and AQUATek 100 delivers:
- Lower maintenance demands for GC inlets, columns and MS sources due to minimized water load.
- Reliable detection of trace VOCs in drinking water without chromatographic artifacts.
- Optimized sample cycle times for higher laboratory productivity.
- Compliance with regulatory requirements for method linearity, detection limits and data quality.
Future Trends and Applications
Emerging directions in VOC water analysis leveraging P&T technology include:
- Integration of advanced sorbent materials for broader compound coverage and enhanced water rejection.
- Automation and remote monitoring to support field-deployable VOC analysis systems.
- Coupling with high-resolution mass spectrometry for non-targeted screening of emerging contaminants.
- Implementation of artificial intelligence for chromatogram interpretation and method optimization.
Conclusion
US EPA Method 524.3 analysis using the Tekmar Lumin P&T concentrator with moisture control and the AQUATek 100 autosampler, coupled to an Agilent 7890B GC and 5977A MSD, achieves accurate and precise quantification of trace VOCs in drinking water. The system demonstrates excellent calibration linearity, low MDLs and robust chromatography with minimal water interference, while enhancing throughput and reducing maintenance.
References
- Munch J.W. Measurement of Purgeable Organic Compounds in Water by Capillary Column Gas Chromatography/Mass Spectrometry; US EPA Method 524.3 - Revision 1.0, June 2009.
Obsah byl automaticky vytvořen z originálního PDF dokumentu pomocí AI a může obsahovat nepřesnosti.
Podobná PDF
US EPA Method 524.4 with the Tekmar Lumin P & T Concentrator and Agilent 7890B GC and 5977A MSD System
2018|Agilent Technologies|Aplikace
US EPA Method 524.4 with the Tekmar Lumin P & T Concentrator and Agilent 7890B GC and 5977A MSD System Application Note Roger Bardsley, Applications Chemist; Teledyne Tekmar P a g e |1 Abstract US EPA Method 524.4 is used…
Klíčová slova
temp, tempbake, bakepurge, purgeether, etherppb, ppblpir, lpirupir, upirlumin, lumindesorb, desorbvariable, variablemcs, mcsprecision, precisionmdl, mdltable, tablebutylbenzene
US EPA Method 524.3 with the Tekmar Lumin P & T Concentrator and Thermo Scientific™ TRACE™ 1310 GC and ISQ™ LT MS System
2017|Thermo Fisher Scientific|Aplikace
US EPA Method 524.3 with the Tekmar Lumin P & T Concentrator and Thermo Scientific™ TRACE™ 1310 GC and ISQ™ LT MS System Application Note Roger Bardsley, Applications Chemist; Teledyne Tekmar P a g e |1 Abstract US EPA Method…
Klíčová slova
temp, temppurge, purgebake, bakeether, etherdesorb, desorblpir, lpirppb, ppbupir, upirlumin, luminvariable, variablemcs, mcsprecision, precisionmdl, mdltime, timetable
US EPA Method 524.4 with the Tekmar Lumin P & T Concentrator and Thermo Scientific™ TRACE™ 1310 GC and ISQ™ LT MS System
2017|Thermo Fisher Scientific|Aplikace
US EPA Method 524.4 with the Tekmar Lumin P & T Concentrator and Thermo Scientific™ TRACE™ 1310 GC and ISQ™ LT MS System Application Note Roger Bardsley, Applications Chemist; Teledyne Tekmar P a g e |1 Abstract US EPA Method…
Klíčová slova
temp, temppurge, purgebake, bakeether, etherdesorb, desorblpir, lpirppb, ppbupir, upirlumin, luminvariable, variablemcs, mcsprecision, precisionmdl, mdltime, timetable
US EPA Method 524.3 with the Teledyne Tekmar Atomx XYZ and the Thermo Scientific™ TRACE™ 1310 GC and ISQ™ LT MS System
2019|Thermo Fisher Scientific|Aplikace
APPLICATION NOTE: TECHNOLOGY: INDUSTRY: AN1907 P+T VOC ENV US EPA Method 524.3 with the Teledyne Tekmar Atomx XYZ and the Thermo Scientific™ TRACE™ 1310 GC and ISQ™ LT MS System Amy Nutter, Applications Chemist; Teledyne Tekmar P a g e…
Klíčová slova
temp, tempbake, bakepurge, purgerinse, rinselpir, lpirupir, upirtert, tertsweep, sweepether, etherxyz, xyzdesorb, desorbatomx, atomxprecision, precisionmrl, mrlaccuracy